E-502 Wiring For gate_capture.exe
Purpose
This program captures two differential analog signals on E-502 while:
DI_SYN1provides the external ADC clockSTART_INprovides the repeating gate- data goes to the PC over the built-in
USB 2.0
The program does not use any digital output.
Analog Inputs
Signal 1 +->X3Signal 1 -->Y3Signal 2 +->X4Signal 2 -->Y4
These pairs correspond to:
ch1:2->X3-Y3ch2:3->X4-Y4
Sync Inputs
2 MHz clockfrom generator ->DI_SYN120 kHz gatefrom generator ->START_IN
Behavior in software:
- every clock edge on
DI_SYN1advances ADC sampling - rising edge of
START_INstarts a capture window - falling edge of
START_INcloses the current window
The stream itself starts immediately after X502_StreamsStart(). Windowing is done in software from the sampled state of START_IN.
Ground
- generator common / reference ground ->
GNDofE-502
Even with differential analog channels, the sync inputs still need a valid common reference with the generator.
USB
E-502 USB 2.0-> PC
ASCII Diagram
Signal source #1 (+) --------------------------> E-502 X3
Signal source #1 (-) --------------------------> E-502 Y3
Signal source #2 (+) --------------------------> E-502 X4
Signal source #2 (-) --------------------------> E-502 Y4
2 MHz generator output ------------------------> E-502 DI_SYN1
20 kHz gate output ----------------------------> E-502 START_IN
Generator common / GND ------------------------> E-502 GND
E-502 USB 2.0 ---------------------------------> PC
Recommended Run Command
.\gate_capture.exe clock:di_syn1_rise clock_hz:2000000 gate_hz:20000 windows:1000 csv:gate_capture.csv svg:gate_capture.svg