# SN9000 (PLANAR SNVNA / Иридиум) Setup
This project controls the PLANAR SN9000 multi-port VNA through the SNVNA
companion application running on an external PC. The production path is:
```text
SN9000 --USB 2.0--> SNVNA host PC --HiSLIP/VISA--> radar_system
```
For complete run-mode instructions see
[`docs/operation_modes.md`](operation_modes.md). For `run_config.json` field
reference see [`docs/run_config.md`](run_config.md).
The SN9000 hardware has no built-in SCPI server; the SNVNA application on the
companion PC exposes the SCPI HiSLIP server (default port `4880`). This
project uses HiSLIP only, with the same `pyvisa` + IVI VISA stack already
required by K209 — there is no additional dependency.
For maximum throughput the driver:
- Uses HiSLIP, not raw TCP Socket.
- Keeps one persistent VISA session.
- Sends `FORM:DATA REAL32` and `FORM:BORD SWAP` (little-endian) once.
- Pre-configures 10 traces covering all S-parameters of the 2×4 matrix so
one trigger drives both stimulus ports.
- Sends the entire acquisition as one synchronized SCPI message:
`TRIG:SING;*OPC?;:SENS:DATA:CORR? S11;:SENS:DATA:CORR? S31;…;:SENS:DATA:CORR? S62`.
The K209 setup notes the same constraint: splitting `TRIG:SING` from the
data queries can return `-211,"Trigger system is not in the trigger wait state"`.
Topology (manual p. 1457):
| Trace | Output position | Stimulus port | Input position | Receiver port |
|-------|-----------------|---------------|----------------|----------------|
| S11 | 0 | 1 | (reflection) | 1 |
| S31 | 0 | 1 | 0 | 3 |
| S41 | 0 | 1 | 1 | 4 |
| S51 | 0 | 1 | 2 | 5 |
| S61 | 0 | 1 | 3 | 6 |
| S22 | 1 | 2 | (reflection) | 2 |
| S32 | 1 | 2 | 0 | 3 |
| S42 | 1 | 2 | 1 | 4 |
| S52 | 1 | 2 | 2 | 5 |
| S62 | 1 | 2 | 3 | 6 |
## Required Components
Install these on the machine that runs the SN9000 smoke test or acquisition
process:
1. SNVNA companion application from Planar.
- The SN9000 hardware is connected to this host over USB 2.0.
2. IVI VISA runtime and development files.
- Must support TCPIP HiSLIP resources.
- Suitable implementations include NI-VISA or Keysight IO Libraries Suite.
- If K209 already works on this host, no additional install is needed.
3. Project Python environment.
- Use the repository virtual environment, not system Python.
- Install `requirements.txt` into `.venv`.
## SNVNA HiSLIP Server
Start SNVNA with the SN9000 connected over USB 2.0. Enable HiSLIP server on
port `4880`. From the SNVNA UI:
```text
System -> Settings -> Remote control network settings -> HiSLIP server -> On
System -> Settings -> Remote control network settings -> HiSLIP port -> 4880
```
Verify that the server is listening:
```bash
ss -ltnp | grep 4880
```
If SNVNA runs on a different machine from `radar_system`, set
`radar.remote_host` to that machine's IP address.
The VISA resource string the driver assembles is:
```text
TCPIP0::<radar.remote_host>::hislip0,<radar.remote_port>::INSTR
```
## `run_config.json`
```json
"radar": {
"model": "sn9000",
"remote_host": "127.0.0.1",
"remote_port": 4880,
"driver_mode": "native"
}
```
The 2×4 virtual switch matrix is enforced automatically; do not edit
`switches.port1` / `switches.port2` or `run.combos` for SN9000 mode — the
config codec rewrites them on load.
## Python Smoke Test
Use the project virtual environment:
```bash
.venv/Scripts/python.exe -m python_app.scripts.sn9000_smoke_test ^
--host 127.0.0.1 --port 4880 ^
--start-hz 1000000 --stop-hz 3000000000 ^
--points 201 --ifbw-hz 10000 --power-dbm -10 ^
--no-preset
```
Expected result:
```text
SN9000 IDN: Planar, SN9000-N, ...
SN9000 collection OK: traces=8, points=201, first_hz=..., last_hz=..., mean_abs_s21=...
```
Use `--no-preset` for the first smoke test to avoid resetting the current
SNVNA session. Remove it when testing the full driver setup path.
## SN9000 Limits
The SNVNA SCPI surface exposes service capability queries identical to K209:
```text
SERV:SWE:FREQ:MAX? Upper frequency bound in Hz.
SERV:SWE:FREQ:MIN? Lower frequency bound in Hz.
SERV:SWE:POIN? Maximum sweep point count.
SERV:SWE:POW:MAX? Upper power bound in dBm.
SERV:SWE:POW:MIN? Lower power bound in dBm.
```
The base SN9000 model covers `0.3 MHz .. 9 GHz`; power range is
`-45 .. +10 dBm` up to 6 GHz, and `-45 .. +2 dBm` from 6 GHz to 9 GHz
(manual p. 58). IF bandwidth selectable in the 1, 1.5, 2, 3, 5, 7 sequence
across decades from `1 Hz` to `300 kHz` (manual p. 58, 1261).